Type N thermocouple

Ni-Cr-Si / Ni-Si · -270 … 1300 °C · IEC 60584-1

At a glance

Conductors (+ / −)
Ni-Cr-Si / Ni-Si
Range of the reference function
-270 … 1300 °C
Sensitivity at 600 °C
38.96 µV/°C
Class 1 tolerance
±1.5 °C or ±0.004 × |t|, whichever is larger, -40 to 1000 °C
Class 2 tolerance
±2.5 °C or ±0.0075 × |t|, whichever is larger, -40 to 1200 °C
Class 3 tolerance
±2.5 °C or ±0.015 × |t|, whichever is larger, -200 to 40 °C
Extension and compensating cable
NX · NCIEC 60584-3: X = extension cable of the same alloys, C = compensating cable of other alloys, matched over a limited temperature range

Wire colours

PositiveNegativeSheath / Jacket
IEC 60584-3 (international)pinkwhitepink
ASTM E230 / ANSI (USA), extension wireorangeredorange

Colours identify the type, not the quality of the cable. Older national codes (e.g. BS 1843, DIN 43714, JIS) used other colours; check before reusing existing cable.

EMF and tolerance at common temperatures (0 °C reference)

Temperature (°C)EMF (mV)Sensitivity (µV/°C)Class 1 (±°C)Class 2 (±°C)Class 3 (±°C)
-200-3.9909.93——±3.00
00.00026.16±1.50±2.50±2.50
1002.77429.64±1.50±2.50—
2005.91332.99±1.50±2.50—
40012.97437.13±1.60±3.00—
60020.61338.96±2.40±4.50—
80028.45539.26±3.20±6.00—
100036.25638.61±4.00±7.50—
120043.84637.19—±9.00—

—: the class is not defined at this temperature. Tolerance of type N at every 100 °C

Full type N EMF tableThermocouple calculator

Typical use and limits

Developed as a more stable alternative to type K: silicon in both legs reduces short-range ordering and oxidation drift, so type N holds its calibration better between 800 and 1200 °C. Output is slightly lower than type K, and it needs instruments and extension cable set up for type N.