Type E thermocouple

Ni-Cr / Cu-Ni · -270 … 1000 °C · IEC 60584-1

At a glance

Conductors (+ / −)
Ni-Cr / Cu-Ni
Range of the reference function
-270 … 1000 °C
Sensitivity at 300 °C
77.91 µV/°C
Class 1 tolerance
±1.5 °C or ±0.004 × |t|, whichever is larger, -40 to 800 °C
Class 2 tolerance
±2.5 °C or ±0.0075 × |t|, whichever is larger, -40 to 900 °C

Wire colours

PositiveNegativeSheath / Jacket
IEC 60584-3 (international)violetwhiteviolet
ASTM E230 / ANSI (USA), extension wirepurpleredpurple

Colours identify the type, not the quality of the cable. Older national codes (e.g. BS 1843, DIN 43714, JIS) used other colours; check before reusing existing cable.

EMF at common temperatures (0 °C reference)

Temperature (°C)EMF (mV)Sensitivity (µV/°C)
-200-8.82525.13
-100-5.23745.17
00.00058.67
1006.31967.52
20013.42174.03
40028.94680.06
60045.09380.66
80061.01778.43

Full type E EMF tableThermocouple calculator

Typical use and limits

The highest output per degree of the standard types, roughly 60 to 80 µV/°C, so it resolves small temperature differences well. Both legs are non-magnetic. Suited to oxidising and inert atmospheres and to sub-zero work.