IEC 60079-29-13-Electrode Cell (WE / RE / CE)

3-Electrode Electrochemical Gas Sensor Potentiostat Circuit

nA–µA Redox Current → Voltage · 3.3 V Ultra-Low-Power Single Rail

A 3-electrode potentiostat holds Reference (RE) at zero current via control op-amp U1 driving Counter (CE), while TIA U2 converts Working (WE) redox current to Vout = Vzero + I_WE·R_tia; a depletion JFET shorts WE–RE when unpowered.

POTENTIOSTAT LOCKED — I_WE = 7.00 µA → V_out = 1.70 V

Zero current flows in RE (< 1 pA); U1 sources 7.00 µA into CE while U2 scales WE to 1.70 V (linear up to 314 ppm).

Normal current flowWarning / ClampFault / OvercurrentFlow speed ∝ current |I|
3-Electrode Cell100 ppmCEREWECO / O₂ / H₂S+−U1 Ctrl1.0 VJFET SHORTI_WE = 7.00 µA−+U2 TIAR_tia = 100 kΩGas ADC Out1.70 V0–3.3 V Linear⚠ TIA RAIL CLIP!

Working Electrode I_WE

7.00 µA

TIA Output Voltage

1.70 V

Linear Full-Scale Range

314 ppm

Reference Electrode I_RE

< 0.001 nA

Shorting JFETOPEN (Active)
RE Control LoopLocked @ 1.0 V
3.3 V ADC Margin1.60 V
ppm

e.g. CO 0–500 ppm (OSHA PEL = 50 ppm).

nA/ppm

Typical Alphasense / Figaro CO cell: 50–100 nA/ppm.

kΩ

Sets V_out = V_zero + I_WE × R_tia (typically 33–200 kΩ).

V

Holds WE and RE at identical virtual potential (0 mV bias for CO).

Presets:

Live Step-by-Step Formulas

Transimpedance Gas Output Equation

V_out = V_zero + (C_ppm × S_nA/ppm) × R_tia = 1.0 V + (100 × 70 nA) × 100 kΩ

⇒ 1.700 V

Dynamic BOM & Component Sizing

RefComponentCalculated RatingCheck
U1 / U2LMP91000 / OPA2333 / ADA4528Low Ib (<100 pA) + Vos < 10 µVOK
Q_JFETMMBFJ177 P-Ch Depletion JFETShorts WE–RE automatically when Vcc=0OK

1. Engineering Overview & How the Circuit Works

Toxic gas detectors (CO, H₂S, NO₂, SO₂, O₂) rely on 3-electrode electrochemical cells (Working WE, Reference RE, Counter CE). A dual-op-amp Potentiostat holds RE at an exact zero-current electrochemical bias via U1 while U2 converts nanoamp-to-microamp oxidation current at WE linearly into voltage.

Why 3 Electrodes Beat 2: Drawing current through the reference electrode polarizes it and ruins calibration. U1 senses RE with < 1 pA input current and drives whatever voltage is needed at CE to supply the full redox reaction current.

Depletion JFET Power-Off Shorting: Leaving WE–RE floating unpowered allows gas/humidity to polarize the cell — requiring up to 24 hours of stabilization upon reboot! A P-channel depletion JFET (MMBFJ177) automatically shorts WE to RE when Vcc = 0 V and opens once powered.

3-Electrode Electrochemical Gas Sensor Potentiostat Circuit — SensorCatalog
Gas Sensor PotentiostatPNG 1200×630

2. Step-by-step: Designing a low-power 3-electrode gas potentiostat

  1. 1Use zero-drift, picoamp-Ib op-amps (OPA2333 / LMP91000)

    Microvolt offset drift and < 100 pA bias keep zero-gas baseline stable over −20…+50 °C.

  2. 2Size R_tia for full-scale ppm inside 0.2–3.1 V ADC window

    Include 1.5× over-range headroom so alarm spikes do not saturate the TIA.

  3. 3Add a 10–33 Ω load resistor R_load in series with WE

    Matches cell datasheet specification and stabilizes the TIA against large cell capacitance (~10–50 µF).

  4. 4Include depletion JFET shorting switch across WE–RE

    Eliminates multi-hour warm-up delays in battery-powered portable detectors.

3. Field Troubleshooting & Multimeter Diagnostics

SymptomRoot CauseMultimeter CheckCorrective Fix
Detector reads 80 ppm in fresh air for hours after battery swapMissing power-off WE–RE shorting JFET allowed cell electrodes to polarize while unpoweredMeasure WE–RE voltage right after powering offInstall MMBFJ177 depletion JFET across WE–RE

4. Frequently Asked Engineering Questions

Do all electrochemical gas sensors use 0 mV bias between WE and RE?

No! CO and H₂S cells typically run at 0 mV bias, while NO, NH₃, and EtO cells require +200 mV to +300 mV bias between WE and RE (set via V_zero divider).