Ultrasonic sensor calculator

Temperature-compensated speed of sound, round-trip echo time and beam spot diameter in air.

In dry air at 20 °C (c = 343.2 m/s), a 3.0 m ultrasonic target takes 17.48 ms round-trip, and a 10° beam spreads to 0.52 m diameter.

Target and air

Result

17.33ms

Echo 17.33 ms · Spot Ø 0.52 m · c = 346.1 m/s

Speed of sound c(T)
346.1 m/s25 °C
Beam spot diameter
0.52 m525 mm
Max ping rate
38.5 Hzwith 1.5× decay guard
Uncomp. 20 °C error
-25.3 mmif no temp sensor

Schematic

Ultrasonic40–200 kHz0.52 mBeam ØToF = 17.33 ms · c(T) = 346.1 m/s
Formulas and standards
c(T) = 331.3 · √(1 + T / 273.15)   t_echo = 2·d / c(T)   D_beam = 2·d·tan(θ/2)

Speed of sound in dry air scales with √T: c(T) = 331.3 · √(1 + T / 273.15) m/s (~0.17%/°C). Beam diameter at target distance d is 2·d·tan(θ/2).

Keep tank walls, ladders and agitator blades outside the full beam cone D_beam to prevent false echoes, and always enable internal temperature compensation outdoors.

Speed of sound and beam geometry

Ultrasonic level and distance sensors measure the round-trip Time of Flight (ToF) of a 40–200 kHz acoustic burst in air. Because speed of sound varies strongly with air temperature (~0.17 %/°C):

  • c(T) = 331.3 · √(1 + T / 273.15) m/s
  • t_echo = 2 · d / c(T)
  • D_beam = 2 · d · tan(θ / 2)

Worked example

At 3.0 m target distance, 20 °C air and a 10° full beam angle:

  • c(20 °C) = 343.2 m/s
  • Round-trip echo time = 17.48 ms (max ping rate ~38 Hz with 1.5× decay guard)
  • Beam spot diameter at target = 0.52 m (keep tank walls > 0.26 m from centerline)

Frequently asked questions

How much error does air temperature cause without compensation?

Air speed of sound changes by roughly 0.17 % per °C (about 0.6 m/s per °C). A 30 °C swing from 20 °C causes roughly 5 % range error (~150 mm at 3 m) unless temperature compensation is enabled.

Why does beam angle matter inside a tank or silo?

Any weld seam, ladder rung, inlet pipe or agitator blade inside the full beam diameter D_beam reflects an early false echo that locks the level reading above the true liquid surface.

Can ultrasonic sensors measure accurately in vacuum or steam?

No — acoustic waves require air/gas molecules and attenuate heavily in dense steam, foam or heavy dust. For vacuum, steam or high-pressure tanks, use 80 GHz FMCW radar or hydrostatic differential pressure.