The numbers on this page are recomputed from the live calculation code every time the site is built. If one of them moved away from its reference, the build would stop and nothing would be published.

Thermocouples against NIST ITS-90

Every value of the NIST SRD 60 tables in 1 °C steps, compared with the reference function used on this site. The tables are rounded to 0.001 mV, so rounding alone can cause up to 0.0005 mV.

TypeRangeTable values comparedLargest deviationSource
Type K thermocouple-270 … 1372 °C16430.00050 mVNIST ITS-90, type K
Type J thermocouple-210 … 1200 °C14110.00050 mVNIST ITS-90, type J
Type T thermocouple-270 … 400 °C6710.00050 mVNIST ITS-90, type T
Type E thermocouple-270 … 1000 °C12710.00050 mVNIST ITS-90, type E
Type N thermocouple-270 … 1300 °C15710.00050 mVNIST ITS-90, type N
Type R thermocouple-50 … 1768.1 °C18190.00050 mVNIST ITS-90, type R
Type S thermocouple-50 … 1768.1 °C18190.00050 mVNIST ITS-90, type S
Type B thermocouple0 … 1820 °C18210.00050 mVNIST ITS-90, type B

PT100 against IEC 60751

Values of the IEC 60751 table (0.01 Ω resolution) next to the Callendar–Van Dusen result.

Temperature (°C)Published (Ω)This site (Ω)Difference (Ω)
-20018.5218.5201+0.0001
-10060.2660.2558−0.0042
0100.00100.0000+0.0000
100138.51138.5055−0.0045
200175.86175.8560−0.0040
300212.05212.0515+0.0015
400247.09247.0920+0.0020
500280.98280.9775−0.0025
600313.71313.7080−0.0020
700345.28345.2835+0.0035
800375.70375.7040+0.0040
850390.48390.4811+0.0011

Nickel RTD against a published DIN 43760 table

311 values of the E+E Elektronik Ni1000 TK6180 table from −60 to 250 °C; largest deviation 0.0050 Ω at a table resolution of 0.01 Ω. Coefficients as printed in the TE Connectivity Ni1000SOT datasheet. Source

Worked reference cases

Each reference value follows from the stated inputs by the formula shown on the tool page and was calculated independently of the site's code.

CalculatorInputsReferenceThis site
4–20 mA converter12 mA · 0 … 10 bar5.0000 bar5.0000 bar
4–20 mA loop power calculator24 V · 10.5 V · 22 mA · 250 Ω · 300 m · 0.75 mm² · 20 °C7.69728 V7.69728 V
NTC thermistor calculator10 kΩ · β 3950 K · 25 °C → 50 °C3588.18 Ω3588.18 Ω
Load cell output calculator2 mV/V · 10 V · 4 × 1000 kg · 1800 kg9.0000 mV9.0000 mV
Pressure unit converter1 bar → psi14.5037738 psi14.5037738 psi
Pressure unit converter1 atm → kPa101.325 kPa101.325 kPa
Pressure transmitter accuracy calculatorURL 100 · 0 … 40 bar · 0.075 % · 0.1 %/10 K × 20 K · 0.1 %/a × 1 a0.22561 bar0.22561 bar
Level to pressure calculator5 m · 998.21 kg/m³ · g 9.80665 m/s²0.489455 bar0.489455 bar
Dew point and humidity calculator20 °C · 50 % RH9.255 °C9.255 °C
Flow rate converter100 US gpm378.5411784 L/min378.5411784 L/min
Vibration converter2.8 mm/s RMS · 50 Hz0.126853 g pk0.126853 g pk
Strain gauge bridge calculatorquarter bridge · GF 2 · 1000 µε · 5 V2.4975 mV2.4975 mV
Strain gauge bridge calculatorfull bridge · GF 2 · 1000 µε · 5 V10.0000 mV10.0000 mV
Temperature measurement error calculatorPT100 class A · 100 °C · 4-wire · 0 … 200 °C · 0.1 K · 0.03 % · 0.1 %/10 K × 20 K0.544151 K0.544151 K
Temperature measurement error calculatortype K class 1 · 500 °C · no other error2.0000 K2.0000 K
ADC resolution calculator16 bit · 0 … 2.5 V38.147 µV38.147 µV
Modbus register decoderFLOAT32 · 0x4120 0x0000 · ABCD10.0000 10.0000
Modbus register decoderFLOAT32 · 0x0FDB 0x4049 · CDAB3.14159274 3.14159274
Sensor accuracy spec converter0 … 100 · reading 50 · 0.1 % rdg + 0.05 % FS · 0.01 %/K × 10 K · 0.1 %/yr0.173205 0.173205
NTC voltage divider calculator10 kΩ / β 3950 at 25 °C · 10 kΩ to 3.3 V · NTC to ground-36.659 mV/K-36.659 mV/K
Anti-aliasing RC filter calculator1 kΩ · 100 nF1591.549 Hz1591.549 Hz
Instrumentation amplifier gain calculatorK 49.4 kΩ · RG 499 Ω99.998 99.998
Current transformer burden calculator5 A · 20 m of 2.5 mm² · 1 VA of devices7.88 VA7.88 VA
Sensor supply voltage drop calculator24 V · 50 mA · 200 m of 0.5 mm²23.312 V23.312 V
Pulse output calculator1024 PPR · ×4 · 3000 rpm204.8 kHz204.8 kHz
Battery life calculatorsleep 5 µA · 10 mA × 100 ms · 40 mA × 50 ms · every 60 s54.9875 µA54.9875 µA

What this does not cover

  • The calculators refuse input that is physically impossible, not input that is merely unusual. Within those limits they calculate exactly what you enter: the choice of values, and the engineering judgement behind them, stay with you.
  • These checks cover the calculation, not your installation: sensor tolerance, wiring, instrument error and site conditions remain yours to assess.
  • The Ex and IP decoders explain a marking. The certificate, the standard and the responsible person decide whether equipment may be used.
  • Found a wrong number? Write to [email protected]. Corrections are checked against the standard and published.