Type T thermocouple

Cu / Cu-Ni · -270 … 400 °C · IEC 60584-1

At a glance

Conductors (+ / −)
Cu / Cu-Ni
Range of the reference function
-270 … 400 °C
Sensitivity at 100 °C
46.78 µV/°C
Class 1 tolerance
±0.5 °C or ±0.004 × |t|, whichever is larger, -40 to 350 °C
Class 2 tolerance
±1 °C or ±0.0075 × |t|, whichever is larger, -40 to 350 °C

Wire colours

PositiveNegativeSheath / Jacket
IEC 60584-3 (international)brownwhitebrown
ASTM E230 / ANSI (USA), extension wireblueredblue

Colours identify the type, not the quality of the cable. Older national codes (e.g. BS 1843, DIN 43714, JIS) used other colours; check before reusing existing cable.

EMF at common temperatures (0 °C reference)

Temperature (°C)EMF (mV)Sensitivity (µV/°C)
-200-5.60315.74
-100-3.37928.39
00.00038.75
502.03642.82
1004.27946.78
2009.28853.15
30014.86258.09
40020.87261.80

Full type T EMF tableThermocouple calculator

Typical use and limits

Copper against constantan: the most accurate base-metal type near ambient and below zero, used in food, pharmaceutical, laboratory and cryogenic work. The copper leg oxidises above about 370 °C, which sets the practical upper limit. Copper conducts heat well, so watch for stem conduction errors in short immersion lengths.