Why vacuum units need care
A vacuum datasheet may quote mbar, Torr, Pa or atm for the same gauge. The conversions are exact, but they are not round numbers: 1 Torr is 133.322 Pa, and 1 mbar is 100 Pa, so 1 mbar is 0.750 Torr. Mixing the two by a few percent is the usual reason two instruments appear to disagree.
The definitions
- 1 atm = 101 325 Pa = 1013.25 mbar = 760 Torr
- 1 Torr = 101 325/760 Pa = 133.322368… Pa
- 1 mbar = 100 Pa
- 1 mTorr = 0.133322368… Pa
The Torr is defined from the atmosphere, not from the millimetre of mercury column, so the conversion is exact and reproducible. The calculator returns all four units from any one of them, with Pa as the reference.
Worked examples
- 1 mbar = 100 Pa = 0.75006 Torr = 750.06 mTorr
- 1 Torr = 133.322 Pa = 1.33322 mbar
- 0.001 mbar = 1e−3 mbar = 0.750 mTorr = 0.1 Pa
- 1 atm = 1013.25 mbar = 760 Torr
A rough scale helps: atmospheric pressure is about 1000 mbar, a rough vacuum pump reaches 1–10 mbar, a good two-stage rotary pump 1e−3 mbar, a turbomolecular pump 1e−7 mbar and an ion-pumped ultra-high vacuum 1e−10 mbar or below.
Gauge types and their ranges
- Pirani and thermocouple gauges: about 1e−3 to 1000 mbar, gas-dependent at the low end.
- Capacitance diaphragm gauges: about 1e−5 to 1000 mbar, with the common industrial version covering 0.1 to 1000 mbar.
- Hot-cathode ion gauges: about 1e−10 to 1e−3 mbar, with a gas-species correction factor.
- Bourdon-tube mechanical gauges: about 1 to 1000 mbar.
The ranges overlap so that a system can be covered by two instruments with a handover point. Check the datasheet for the exact model and the calibration gas: a Pirani gauge calibrated for nitrogen reads differently in helium or hydrogen.
Where the units meet
The same vacuum appears in other calculators: the pressure converter covers bar, psi, kPa and mmH₂O, and the dew-point calculator uses pressure dew point for compressed air, where the pressure is usually given in bar but the humidity in mbar of vapour pressure.
Common mistakes
Rounding the Torr factor. Using 133 Pa instead of 133.322 Pa is a 0.24 % error; at 1e−3 mbar it is small, at 760 Torr it is 245 Pa.
Mixing gauge and absolute pressure. A vacuum gauge may read absolute pressure or the difference from atmosphere; the datasheet says which one.
Assuming a gauge reads to zero. Every gauge has a lower limit set by its principle; below that the reading is noise.
Comparing readings in different units without converting. 1 mbar and 1 Torr differ by 25 %.
Frequently asked questions
How many Torr is 1 mbar?
0.75006 Torr. The exact factor is 100 × 760/101325.
Is 1 Torr exactly 1 mmHg?
Historically close, but the Torr is defined as 1/760 atm, while 1 mmHg depends on the density of mercury and gravity. They differ by about 1e−7 relative.
What is a rough vacuum in mbar?
Rough vacuum is usually taken as 1 to 1000 mbar, fine vacuum 1e−3 to 1 mbar and high vacuum below 1e−3 mbar; the boundaries vary by industry.
Can I convert a gauge reading below its range?
The unit conversion is still valid, but the number is outside the gauge’s specification. Below the range the reading is not a measurement.