pH Nernst mV & conductivity/TDS calculator

Temperature-compensated pH electrode Nernst EMF (mV) and 25 °C EC / TDS water quality conversion.

At 25 °C the Nernst slope is −59.16 mV/pH (pH 4.00 = +177.5 mV, ASTM D1293), and 1413 µS/cm water conductivity corresponds to roughly 707 ppm TDS (k = 0.50).

Solution parameters

Result

+177.5mV

pH 4.00 = +177.5 mV · TDS ≈ 707 ppm

Nernst slope S(T)
-59.16 mV/pH25 °C (ATC)
Compensated EC₂₅
1413.0 µS/cmreferenced to 25 °C (2%/°C)
Total dissolved solids
706.5 ppmk = 0.50
Water resistivity
0.71 kΩ·cm1000 / EC₂₅

Schematic

pH 0 (+414 mV)pH 7 (0 mV)pH 14 (−414 mV)+177.5 mVpH 4.00EC₂₅ / TDS1413 µS/cm707 ppm TDS
Formulas and standards
E = −59.16 · ((T + 273.15) / 298.15) · (pH − 7) mV   EC₂₅ = EC / (1 + 0.02·(T − 25))

Nernst potential is E = −59.16 · ((T + 273.15)/298.15) · (pH − 7) mV (ASTM D1293). Conductivity compensates to 25 °C using α = 2.0 %/°C.

Because pH glass bulbs have 100–500 MΩ impedance, always buffer with a femtoamp op-amp (LMP7721) and driven Guard Ring.

pH Nernst potential and 25 °C conductivity compensation

A pH glass electrode obeys the temperature-dependent Nernst equation (ASTM D1293):

  • S(T) = −59.16 · ((T + 273.15) / 298.15) mV/pH
  • E = (pH − 7.0) · S(T) mV

Water electrical conductivity (EC) rises roughly +2.0 % per °C, so instruments normalize readings to 25 °C: EC₂₅ = EC / (1 + 0.02 · (T − 25)), and estimate TDS = EC₂₅ · k.

Worked example

At 25 °C, pH 4.00 buffer produces +177.5 mV. A 1413 µS/cm standard solution at 25 °C equals 706.5 ppm TDS (k = 0.50) and 0.71 kΩ·cm resistivity.

Frequently asked questions

Why does pH require automatic temperature compensation (ATC)?

The Nernst slope shifts from −54.20 mV/pH at 0 °C to −64.12 mV/pH at 50 °C. Without ATC, any solution away from pH 7.00 drifts significantly with water temperature.

Why does a normal op-amp ruin pH glass electrode accuracy?

Glass electrodes have 100–500 MΩ internal resistance, so even 1 nA of bias current causes a 500 mV (>8 pH) error — see our pH glass electrode & Guard Ring circuit.

What TDS factor k should I use?

Use k = 0.50 for standard NaCl / tap water, k ≈ 0.64 for potassium chloride (KCl) calibration standards, and k ≈ 0.70 for natural mineral-rich water (442 standard).