Converting gas sensor readings
Gas detectors measure volume fraction (ppm, %Vol, %LEL), whereas environmental and occupational exposure limits (OSHA, NIOSH, EU OEL) are frequently specified in mass concentration (mg/m³). Because gas density depends on temperature and pressure, the conversion factor changes with the molar volume:
- V_m(T, P) = 22.414 · ((T + 273.15) / 273.15) · (101.325 / P) L/mol
- mg/m³ = ppm · (M / V_m)
Worked example
For carbon monoxide (CO, M = 28.01 g/mol) at 25 °C and 101.325 kPa:
- V_m = 24.465 L/mol
- 50 ppm × (28.01 / 24.465) = 57.24 mg/m³
- Volume fraction = 50 / 10 000 = 0.0050 %Vol
- With LEL = 10.9 %Vol (IEC 60079-20-1), 50 ppm is 0.046 %LEL
Frequently asked questions
Why does ppm to mg/m³ change with temperature?
Because ppm is a volume-to-volume ratio while mg/m³ is mass per cubic metre. Heating air expands the gas so fewer moles occupy one cubic metre, lowering mg/m³ for the same ppm reading.
What is the difference between 24.45 and 22.41 L/mol?
22.414 L/mol is the molar volume at STP (0 °C, 101.325 kPa); 24.465 L/mol is the molar volume at 25 °C and 101.325 kPa, commonly used in occupational hygiene tables.
How does %LEL relate to %Vol?
100 %LEL equals the lower explosive limit of that specific gas in air (for example 4.4 %Vol for methane CH₄ or 4.0 %Vol for H₂S per IEC 60079-20-1).
Which sensor front-end is used for electrochemical CO and H₂S cells?
Three-electrode electrochemical gas cells use a dual op-amp potentiostat with a depletion JFET shorting switch when unpowered — see the 3-electrode gas potentiostat circuit.