SensorCatalog

Gaskonzentrations-Rechner

Umrechnung von Gassensorwerten zwischen ppm, mg/m³, %Vol und %UEG mit T- und P-Korrektur.

Bei 25 °C und 101,325 kPa (Vm = 24,47 L/mol) entsprechen 50 ppm CO (28,01 g/mol) genau 57,24 mg/m³, 0,0050 %Vol bzw. 0,046 %UEG (IEC 60079-20-1).

Gas und Bedingungen

Ergebnis

57,24mg/m³

CO: 50 ppm = 57,24 mg/m³

Volumenanteil
0,0050 %1 %Vol = 10 000 ppm
Explosionsgrenze
0,05 % LEL100% LEL = 10,9 %Vol
Molare Masse
28,01 g/molCO
Molvolumen
24,47 L/mol25 °C · 101,33 kPa

Prinzipskizze

Volume Conc.50,0 ppmCOM / V_m(T,P)Mass Conc. & LEL57,24 mg/m³0,05 % LEL
Formeln und Normen
V_m = 22.414 · ((T + 273.15) / 273.15) · (101.325 / P)   mg/m³ = ppm · M / V_m

Das Molvolumen wird per idealem Gasgesetz ab Normzustand (22,414 L/mol) skaliert; UEG-Werte nach IEC 60079-20-1.

Elektrochemische Sensoren kalibrieren in Volumenanteilen (ppm / %UEG); Arbeitsplatzgrenzwerte werden oft in mg/m³ angegeben.

Converting gas sensor readings

Gas detectors measure volume fraction (ppm, %Vol, %LEL), whereas environmental and occupational exposure limits (OSHA, NIOSH, EU OEL) are frequently specified in mass concentration (mg/m³). Because gas density depends on temperature and pressure, the conversion factor changes with the molar volume:

  • V_m(T, P) = 22,414 · ((T + 273,15) / 273,15) · (101,325 / P) L/mol
  • mg/m³ = ppm · (M / V_m)

Worked example

For carbon monoxide (CO, M = 28,01 g/mol) at 25 °C and 101,325 kPa:

  • V_m = 24,465 L/mol
  • 50 ppm × (28,01 / 24,465) = 57,24 mg/m³
  • Volume fraction = 50 / 10 000 = 0,0050 %Vol
  • With LEL = 10,9 %Vol (IEC 60079-20-1), 50 ppm is 0,046 %LEL

Frequently asked questions

Why does ppm to mg/m³ change with temperature?

Because ppm is a volume-to-volume ratio while mg/m³ is mass per cubic metre. Heating air expands the gas so fewer moles occupy one cubic metre, lowering mg/m³ for the same ppm reading.

What is the difference between 24,45 and 22,41 L/mol?

22,414 L/mol is the molar volume at STP (0 °C, 101,325 kPa); 24,465 L/mol is the molar volume at 25 °C and 101,325 kPa, commonly used in occupational hygiene tables.

How does %LEL relate to %Vol?

100 %LEL equals the lower explosive limit of that specific gas in air (for example 4,4 %Vol for methane CH₄ or 4,0 %Vol for H₂S per IEC 60079-20-1).

Which sensor front-end is used for electrochemical CO and H₂S cells?

Three-electrode electrochemical gas cells use a dual op-amp potentiostat with a depletion JFET shorting switch when unpowered — see the 3-electrode gas potentiostat circuit.