温度・圧力によるモル体積補正付き ppm / mg/m³ / %Vol / %LEL 相互換算。

25 °C・101.325 kPa(Vm = 24.47 L/mol)で CO 50 ppm は 57.24 mg/m³・0.0050 %Vol・0.046 %LEL です(IEC 60079-20-1)。

ガスと条件

結果

57.24mg/m³

CO: 50 ppm = 57.24 mg/m³

体積率
0.0050 %1 %Vol = 10 000 ppm
爆発下限界比
0.05 % LEL100% LEL = 10.9 %Vol
分子量
28.01 g/molCO
モル体積
24.47 L/mol25 °C · 101.33 kPa

概略図

Volume Conc.50.0 ppmCOM / V_m(T,P)Mass Conc. & LEL57.24 mg/m³0.05 % LEL
計算式と規格
V_m = 22.414 · ((T + 273.15) / 273.15) · (101.325 / P)   mg/m³ = ppm · M / V_m

標準状態(0 °C・101.325 kPa で 22.414 L/mol)から理想気体の法則で補正。LEL は IEC 60079-20-1 準拠。

電気化学式センサは体積比(ppm / %LEL)で校正され、環境・作業環境基準は mg/m³ で規定されることが多くあります。

Converting gas sensor readings

Gas detectors measure volume fraction (ppm, %Vol, %LEL), whereas environmental and occupational exposure limits (OSHA, NIOSH, EU OEL) are frequently specified in mass concentration (mg/m³). Because gas density depends on temperature and pressure, the conversion factor changes with the molar volume:

  • V_m(T, P) = 22.414 · ((T + 273.15) / 273.15) · (101.325 / P) L/mol
  • mg/m³ = ppm · (M / V_m)

Worked example

For carbon monoxide (CO, M = 28.01 g/mol) at 25 °C and 101.325 kPa:

  • V_m = 24.465 L/mol
  • 50 ppm × (28.01 / 24.465) = 57.24 mg/m³
  • Volume fraction = 50 / 10 000 = 0.0050 %Vol
  • With LEL = 10.9 %Vol (IEC 60079-20-1), 50 ppm is 0.046 %LEL

Frequently asked questions

Why does ppm to mg/m³ change with temperature?

Because ppm is a volume-to-volume ratio while mg/m³ is mass per cubic metre. Heating air expands the gas so fewer moles occupy one cubic metre, lowering mg/m³ for the same ppm reading.

What is the difference between 24.45 and 22.41 L/mol?

22.414 L/mol is the molar volume at STP (0 °C, 101.325 kPa); 24.465 L/mol is the molar volume at 25 °C and 101.325 kPa, commonly used in occupational hygiene tables.

How does %LEL relate to %Vol?

100 %LEL equals the lower explosive limit of that specific gas in air (for example 4.4 %Vol for methane CH₄ or 4.0 %Vol for H₂S per IEC 60079-20-1).

Which sensor front-end is used for electrochemical CO and H₂S cells?

Three-electrode electrochemical gas cells use a dual op-amp potentiostat with a depletion JFET shorting switch when unpowered — see the 3-electrode gas potentiostat circuit.