pH Nernst-mV & Leitfähigkeit/TDS-Rechner

Temperaturkompensierte pH-Nernst-EMK (mV) und 25-°C-Leitfähigkeit / TDS-Umrechnung.

Bei 25 °C liegt die Nernst-Steigung bei −59,16 mV/pH (pH 4,00 = +177,5 mV, ASTM D1293); 1413 µS/cm entsprechen ca. 707 ppm TDS (k = 0,50).

Lösungsparameter

Ergebnis

+177,5mV

pH 4,00 = +177,5 mV · TDS ≈ 707 ppm

Nernst-Steigung S(T)
-59,16 mV/pH25 °C (ATC)
Kompensierte EC₂₅
1413,0 µS/cmbezogen auf 25 °C (2 %/°C)
TDS-Gehalt
706,5 ppmk = 0,50
Spez. Widerstand
0,71 kΩ·cm1000 / EC₂₅

Prinzipskizze

pH 0 (+414 mV)pH 7 (0 mV)pH 14 (−414 mV)+177,5 mVpH 4,00EC₂₅ / TDS1413 µS/cm707 ppm TDS
Formeln und Normen
E = −59.16 · ((T + 273.15) / 298.15) · (pH − 7) mV   EC₂₅ = EC / (1 + 0.02·(T − 25))

Nernst: E = −59,16 · ((T + 273,15)/298,15) · (pH − 7) mV (ASTM D1293). EC₂₅ mit α = 2,0 %/°C.

pH-Glaselektroden haben 100–500 MΩ Quellwiderstand — immer mit Femtoampere-Puffer (LMP7721) und Guard Ring betreiben.

pH Nernst potential and 25 °C conductivity compensation

A pH glass electrode obeys the temperature-dependent Nernst equation (ASTM D1293):

  • S(T) = −59,16 · ((T + 273,15) / 298,15) mV/pH
  • E = (pH − 7,0) · S(T) mV

Water electrical conductivity (EC) rises roughly +2,0 % per °C, so instruments normalize readings to 25 °C: EC₂₅ = EC / (1 + 0,02 · (T − 25)), and estimate TDS = EC₂₅ · k.

Worked example

At 25 °C, pH 4,00 buffer produces +177,5 mV. A 1413 µS/cm standard solution at 25 °C equals 706,5 ppm TDS (k = 0,50) and 0,71 kΩ·cm resistivity.

Frequently asked questions

Why does pH require automatic temperature compensation (ATC)?

The Nernst slope shifts from −54,20 mV/pH at 0 °C to −64,12 mV/pH at 50 °C. Without ATC, any solution away from pH 7,00 drifts significantly with water temperature.

Why does a normal op-amp ruin pH glass electrode accuracy?

Glass electrodes have 100–500 MΩ internal resistance, so even 1 nA of bias current causes a 500 mV (>8 pH) error — see our pH glass electrode & Guard Ring circuit.

What TDS factor k should I use?

Use k = 0,50 for standard NaCl / tap water, k ≈ 0,64 for potassium chloride (KCl) calibration standards, and k ≈ 0,70 for natural mineral-rich water (442 standard).