熱電対計算ツール
NIST ITS-90 に基づく温度と計器の指示電圧(冷接点補償付き)。どちらの欄にも入力できます。
熱電対
種類
範囲-270 °C~1372 °C
変換
結果
- 熱起電力(基準 0 °C)
- 20.6443 mV指示値 + 冷接点の熱起電力
- 冷接点の熱起電力
- 1.0002 mV25 °C 時
- 感度
- 42.63 µV/°Cゼーベック係数
- クラス 1 許容差
- ±2.0 °Cクラス 2:±3.8 °C
表
| 温度 (°C) | 熱起電力、基準 0 °C(mV) | 感度(µV/°C) |
|---|---|---|
| 0 | 0.000 | 39.45 |
| 10 | 0.397 | 39.91 |
| 20 | 0.798 | 40.33 |
| 30 | 1.203 | 40.69 |
| 40 | 1.612 | 41.00 |
| 50 | 2.023 | 41.25 |
| 60 | 2.436 | 41.42 |
| 70 | 2.851 | 41.52 |
| 80 | 3.267 | 41.54 |
| 90 | 3.682 | 41.49 |
| 100 | 4.096 | 41.37 |
| 110 | 4.509 | 41.19 |
| 120 | 4.920 | 40.97 |
| 130 | 5.328 | 40.73 |
| 140 | 5.735 | 40.49 |
| 150 | 6.138 | 40.28 |
| 160 | 6.540 | 40.10 |
| 170 | 6.941 | 39.98 |
| 180 | 7.340 | 39.92 |
| 190 | 7.739 | 39.91 |
| 200 | 8.138 | 39.97 |
| 210 | 8.539 | 40.06 |
| 220 | 8.940 | 40.20 |
| 230 | 9.343 | 40.36 |
| 240 | 9.747 | 40.53 |
| 250 | 10.153 | 40.71 |
| 260 | 10.561 | 40.88 |
| 270 | 10.971 | 41.04 |
| 280 | 11.382 | 41.19 |
| 290 | 11.795 | 41.32 |
| 300 | 12.209 | 41.45 |
| 310 | 12.624 | 41.55 |
| 320 | 13.040 | 41.65 |
| 330 | 13.457 | 41.74 |
| 340 | 13.874 | 41.83 |
| 350 | 14.293 | 41.91 |
| 360 | 14.713 | 41.98 |
| 370 | 15.133 | 42.05 |
| 380 | 15.554 | 42.12 |
| 390 | 15.975 | 42.18 |
| 400 | 16.397 | 42.24 |
| 410 | 16.820 | 42.30 |
| 420 | 17.243 | 42.35 |
| 430 | 17.667 | 42.40 |
| 440 | 18.091 | 42.45 |
| 450 | 18.516 | 42.49 |
| 460 | 18.941 | 42.53 |
| 470 | 19.366 | 42.56 |
| 480 | 19.792 | 42.59 |
| 490 | 20.218 | 42.61 |
| 500 | 20.644 | 42.63 |
| 510 | 21.071 | 42.64 |
| 520 | 21.497 | 42.65 |
| 530 | 21.924 | 42.65 |
| 540 | 22.350 | 42.64 |
| 550 | 22.776 | 42.63 |
| 560 | 23.203 | 42.62 |
| 570 | 23.629 | 42.60 |
| 580 | 24.055 | 42.57 |
| 590 | 24.480 | 42.54 |
| 600 | 24.905 | 42.51 |
| 610 | 25.330 | 42.46 |
| 620 | 25.755 | 42.42 |
| 630 | 26.179 | 42.37 |
| 640 | 26.602 | 42.31 |
| 650 | 27.025 | 42.25 |
| 660 | 27.447 | 42.19 |
| 670 | 27.869 | 42.12 |
| 680 | 28.289 | 42.05 |
| 690 | 28.710 | 41.98 |
| 700 | 29.129 | 41.90 |
| 710 | 29.548 | 41.82 |
| 720 | 29.965 | 41.73 |
| 730 | 30.382 | 41.65 |
| 740 | 30.798 | 41.56 |
| 750 | 31.213 | 41.47 |
| 760 | 31.628 | 41.38 |
| 770 | 32.041 | 41.29 |
| 780 | 32.453 | 41.19 |
| 790 | 32.865 | 41.10 |
| 800 | 33.275 | 41.00 |
| 810 | 33.685 | 40.90 |
| 820 | 34.093 | 40.80 |
| 830 | 34.501 | 40.71 |
| 840 | 34.908 | 40.61 |
| 850 | 35.313 | 40.51 |
| 860 | 35.718 | 40.41 |
| 870 | 36.121 | 40.31 |
| 880 | 36.524 | 40.21 |
| 890 | 36.925 | 40.11 |
| 900 | 37.326 | 40.00 |
| 910 | 37.725 | 39.90 |
| 920 | 38.124 | 39.80 |
| 930 | 38.522 | 39.70 |
| 940 | 38.918 | 39.60 |
| 950 | 39.314 | 39.50 |
| 960 | 39.708 | 39.40 |
| 970 | 40.101 | 39.29 |
| 980 | 40.494 | 39.19 |
| 990 | 40.885 | 39.09 |
| 1000 | 41.276 | 38.98 |
行をクリックすると変換欄に読み込みます。
計算式と規格
熱起電力は NIST ITS-90 基準関数(NIST Monograph 175、IEC 60584-1 の基礎)によるもので、0 °C 基準です。K 熱電対は 0 °C 以上で指数項を含みます。
E(Thot, 0 °C) = Emeter + E(Tcj, 0 °C)
温度は基準関数を数値的に逆算して求めるため、別個の逆関数多項式による小さな誤差を避けられます。冷接点を 0 にすると規準熱起電力表の値になります。
熱電対プローブ、MI ケーブル、ヘッド取付形変換器をお探しですか?
SensorSpan で熱電対を見る