IEC 607512-wire / 3-wire / 4-wire Kelvin

Excitation RTD PT100 / PT1000 en 2, 3 et 4 Fils (IEC 60751)

IEC 60751 Constant-Current Sense · 0.1–2.0 mA Ratiometric IDAC

La PT100 ne varie que de 0,385 Ω/°C : le câblage 2 fils ajoute une forte erreur +2·Rlead. Le 3 fils compense les fils égaux et le 4 fils Kelvin annule totalement la résistance de ligne.

PT100 (4-WIRE) = 138.51 Ω · Lead Error -0.000 °C · V_RTD = 138.5 mV

IEC 60751 exact measurement at 100 °C. Lead resistance (1.5 Ω/wire) is cancelled and self-heating is only +0.055 °C.

Courant normalAlerte / ÉcrêtageDéfaut / SurintensitéVitesse ∝ |I|
Platinum RTDIEC 60751PT100 138.5ΩLead: 1.5 Ω/wire⚠ OPEN RTD ELEMENT!I_exc = 1.0 mA4-wire sense ≈ 0 nA (zero drop) · 2-wire adds +2·Rlead!MAX31865 / ADS124S08Ratiometric RTD ADC100.0 °C138.5 mVErr +-0.00°CSeries Rref cancels IDAC drift0 V (AGND)

Measured Temperature

100.00 °C

Lead-Wire Temp Error

+-0.000 °C

ADC Sense Voltage

138.5 mV

RTD Self-Heating

+0.055 °C

Lead Cancellation Mode4-WIRE KELVIN (100% Cancel)
IEC 60751 Element R(T)PT100: 138.51 Ω (0.38 Ω/°C)
Element Self-Heating Power0.139 mW
Ratiometric Reference Rref400–430 Ω (±0.1%)
°C

IEC 60751 Callendar–Van Dusen exact Pt curve (−100 to +500 °C).

Ω

Set 100 Ω for PT100 or 1000 Ω for PT1000 (10× less lead-wire error!).

mA

Keep ≤ 1.0 mA for PT100 (≤ 0.25 mA for PT1000) to avoid self-heating!

Ω

20 m of 0.34 mm² wire ≈ 1.0 Ω per wire; long 100 m cable ≈ 5 Ω.

Préréglages :

Formules pas à pas en direct

1. IEC 60751 Callendar–Van Dusen resistance

R(T) = R0 · (1 + A·T + B·T²) = PT100 at 100 °C

⇒ 138.505 Ω (dR/dT = 0.379 Ω/°C)

2. Lead-wire temperature error

ΔT = 0 (Kelvin sense current ≈ 0) = 0.000 Ω / 0.379 Ω/°C

⇒ +-0.000 °C

3. Self-heating power in element

P_RTD = I_exc² · R(T) = (1.0 mA)² × 138.5 Ω

⇒ 0.139 mW (+0.055 °C in air)

Nomenclature dynamique (BOM)

Réf.ComposantCalibrage calculéÉtat
U1Ratiometric 24-bit RTD ADC / ConverterMAX31865 (15-bit) or ADS124S08 (24-bit dual-IDAC)OK
RrefUltra-Low-Drift Reference Resistor402 Ω ±0.05%, ≤10 ppm/°COK
EMIMatched RC Input EMI Filter1 kΩ + 100 nF diff / 10 nF CM (C0G)OK

1. Présentation & principe de fonctionnement

Platinum Resistance Temperature Detectors (PT100 and PT1000 per IEC 60751) deliver unmatched long-term stability (±0.15 °C Class A) from −200 °C to +850 °C — provided the excitation circuit cancels lead-wire resistance and limits self-heating.

Because PT100 changes by only 0.385 Ω per °C, even 1.5 Ω of lead resistance per conductor adds a catastrophic +7.8 °C offset in a 2-wire connection! Understanding 2-wire vs 3-wire dual-IDAC vs 4-wire Kelvin excitation is mandatory for every instrumentation engineer.

4-Wire Kelvin (Gold Standard): Two outer wires force constant current Iexc through the platinum element while two inner high-impedance wires connect straight to the differential ADC (drawing nanoamps). Since I_sense ≈ 0, lead resistance produces ZERO voltage drop — even if all four wires have different resistances!

3-Wire Dual-IDAC (Industrial Workhorse): Two matched constant-current sources (IDAC1 = IDAC2) flow through two identical lead wires, cancelling R_lead1 − R_lead2 = 0 on one fewer copper conductor than 4-wire. Ratiometric wiring through a single precision Rref cancels IDAC drift automatically.

Excitation RTD PT100 / PT1000 en 2, 3 et 4 Fils (IEC 60751) — SensorCatalog
Excitation RTD 2/3/4 FilsPNG 1200×630

2. Step-by-step: Wiring and sizing a ratiometric PT100 / PT1000 excitation circuit

  1. 1Choose 3-wire or 4-wire for PT100 (never 2-wire beyond 1 m!)

    For PT100 over any run > 1 m, specify 3-wire (industrial standard) or 4-wire (lab/calibration). If stuck with 2-wire cabling, switching from PT100 to PT1000 cuts lead-wire temperature error by 10× immediately.

  2. 2Keep excitation current ≤ 1.0 mA (PT100) or ≤ 0.25 mA (PT1000)

    Calculate P = Iexc² × R(T). Keeping dissipation under 0.25 mW ensures self-heating stays below 0.1 °C even in still air (0.5–1.0 mA for PT100; 0.1–0.25 mA for PT1000).

  3. 3Use a single low-drift Rref in series for true ratiometric ADC conversion

    Route the same excitation current through Rref (402 Ω for PT100, 4.02 kΩ for PT1000, ±0.05%, ≤10 ppm/°C) to generate the ADC’s Vref — making Iexc accuracy and drift cancel out completely!

  4. 4Match RC filter resistors symmetrically and verify open/short detection

    Keep input filter resistors identical (e.g. 1 kΩ ±0.1%) with C_diff = 100 nF and C_cm = 10 nF C0G, and test open-lead burnout reporting during commissioning.

3. Dépannage terrain & mesures au multimètre

SymptômeCause racineContrôle multimètreAction corrective
PT100 reads consistently +3 to +10 °C higher than a calibrated referenceWired as 2-wire (or 3-wire jumper installed at PLC end instead of sensor end!)Measure resistance between the two same-colored leads (e.g. Red–Red): must be < 3 ΩRun true 3rd wire all the way to the RTD probe head, or upgrade to 4-wire
3-wire PT100 still has a ±0.5–1.0 °C offset after wiring properlyLead-wire resistance mismatch (oxidized terminal block screw on one wire) or IDAC mismatchSwap the two identical-color leads or enable IDAC chopping in ADS124S08Tighten/clean terminals, use matched 3-core cable, or switch to 4-wire Kelvin

4. Questions fréquentes (FAQ)

Why is 4-wire RTD more accurate than 3-wire RTD?

3-wire assumes two lead wires have identical resistance — any oxidized terminal or contact resistance difference creates an uncancelled error. 4-wire Kelvin draws virtually zero current on the sense leads, cancelling lead resistance completely even when wires don’t match.

How do I identify the wires on a standard IEC 60751 3-wire or 4-wire PT100?

Per IEC 60751: 3-wire has 2 Red + 1 White (the two Red leads connect to the same side of the element; Red–Red measures just loop lead resistance ~1 Ω, Red–White measures ~100 Ω at 0 °C). 4-wire has 2 Red + 2 White (one Red+White pair on each side).

Why should PT1000 use lower excitation current than PT100?

Self-heating power is P = I² × R. At the same 1 mA current, a PT1000 dissipates 10× more heat than a PT100 (1.0 mW vs 0.1 mW!), warming itself by ~0.4 °C in air. Dropping Iexc to 0.2 mA cuts PT1000 heat 25× while still producing a clean 200 mV signal!

What is the purpose of Rref in a MAX31865 or ADS124S08 RTD circuit?

Rref turns the circuit into a ratiometric resistance bridge: the ADC measures the ratio V_RTD / V_Rref = R_RTD / R_ref directly, so excitation current drift disappears completely. Only Rref’s own stability matters (use ≤10 ppm/°C).

Can I jumper a 2-wire PT100 to a 3-wire PLC input card?

Yes, jumper the third compensation terminal to one of the two wires at the PLC block so the card won’t fault — BUT understand this provides ZERO lead-wire cancellation (cable resistance still adds full +2·Rlead error).

Why does an RTD read wildly negative or max-scale when a single wire breaks?

With the current path broken, the excitation IDAC drives the input node straight up to its compliance rail (+3.3 V), which exceeds the max valid resistance code and triggers the ADC/PLC Open-RTD fault flag.