1. Présentation & principe de fonctionnement
A pH glass electrode is one of the highest-impedance sensors in engineering: 100 MΩ to 1 GΩ source resistance generating ±414 mV via the Nernst equation (−59.16 mV/pH at 25 °C). Measuring it accurately requires a femtoamp-bias op-amp (LMP7721 / LMC6001), a driven Guard Ring around the input pin, and a precision mid-rail Vref shift.
Why Ordinary Op-Amps Fail Catastrophically: Even a "good" 1 nA input-bias op-amp flowing through a 500 MΩ glass bulb creates 1 nA × 500 MΩ = 500 mV error — more than 8 full pH units! Only fA-class amplifiers (Ib = 3 fA → 1.5 µV error) preserve ±0.01 pH accuracy.
Driven Guard Ring Physics: Surrounding the high-Z BNC center pin with a bare copper ring driven by the unity-gain buffer output keeps the potential difference across the FR4 surface at 0.000 V, reducing PCB moisture leakage current to zero!

2. Step-by-step: Building a ±0.01 pH glass-electrode front-end
1Select a femtoamp-class unity buffer (LMP7721 / LMC6001)
LMP7721 provides dedicated empty guard pins right next to the input pin for effortless PCB guard routing.
2Bias the reference electrode to a clean 1.65 V / 2.048 V Vref
Shifts the ±414 mV Nernst swing cleanly into 1.2–2.1 V on a single 3.3 V rail.
3Route an exposed-copper driven Guard Ring on top and bottom layers
Remove solder mask over the guard ring and ultrasonically wash all rosin flux residue.
4Measure solution temperature (PT1000/NTC) for Nernst slope ATC
Slope shifts from −54.2 mV/pH at 0 °C to −64.1 mV/pH at 50 °C — ATC is mandatory outside 25 °C.
3. Dépannage terrain & mesures au multimètre
| Symptôme | Cause racine | Contrôle multimètre | Action corrective |
|---|---|---|---|
| pH reading jumps wildly whenever a hand approaches the board | Unshielded high-impedance BNC/trace picking up 50/60 Hz capacitive body hum | Place grounded metal shield over front-end and check stability | Use PTFE coaxial BNC, driven Guard Ring, and metal EMI shield can |
4. Questions fréquentes (FAQ)
Why is the pH electrode voltage 0 mV at pH 7.00?
Because the internal reference buffer inside the glass bulb is formulated at pH 7.00, so zero hydrogen-ion concentration gradient exists across the glass membrane at neutral pH 7.
5. Calculateurs associés, guides et autres circuits
- Calculateur pH Nernst mV & conductivité/TDS
- Calculateur de résolution d’un CAN
- Calculateur d’erreur de mesure de température
- Pont Diviseur NTC + Suiveur AOP & Filtre RC
- Pont de Wheatstone + Amplificateur d’Instrumentation (INA)
- Excitation RTD PT100 / PT1000 en 2, 3 et 4 Fils (IEC 60751)
- Frontal Thermocouple : CJC, Filtre EMI & Rupture (Burnout)
- Sortie Capteur PNP 3 Fils (Source) — Schéma & Simulateur
- Sortie Capteur NPN 3 Fils (Sink / Collecteur Ouvert)
- Circuit NAMUR 2 Fils Sécurité Intrinsèque (IEC 60947-5-6)
- Boucle de Courant 4–20 mA 2 Fils & Récepteur Protégé