1. Überblick & Funktionsprinzip
The Wheatstone bridge paired with a precision Instrumentation Amplifier (INA128 / INA333 / AD620) is the cornerstone circuit for load cells, torque sensors, foil strain gauges, and piezoresistive pressure dies.
A 2.0 mV/V load cell excited at 5 V outputs only 10.0 mV at full rated load sitting on top of a 2.50 V common-mode voltage. An instrumentation amplifier strips away the 2.50 V common-mode (>100 dB CMRR) and amplifies the tiny millivolt signal by G = 1 + 50 kΩ / Rg with a single precision gain resistor.
Quarter vs Full Bridge: A single active foil gauge (quarter bridge) outputs ΔVin ≈ Vexc·(ΔR/R0)/4 and shifts with ambient temperature. A full 4-arm load cell stretches two opposite arms (+ΔR) while compressing the other two (−ΔR), yielding 4× higher sensitivity, exact linearity, and automatic first-order temperature cancellation.
The Diamond Plot & Low-Z Vref Rule: Every instrumentation amplifier has an internal input/output common-mode "diamond" boundary. Keeping bridge common-mode centered at Vexc/2 maximizes swing, and driving the INA REF pin with a low-impedance op-amp buffer (NEVER a bare resistor divider!) preserves its >100 dB CMRR.

2. Step-by-step: Designing a precision Wheatstone bridge + In-Amp circuit
1Calculate max bridge differential ΔVin(max) = Vexc × sensitivity
Multiply excitation voltage by rated sensitivity (e.g. 5 V × 2 mV/V = 10 mV) and add 20% headroom for tare weight and zero-balance offset before choosing gain.
2Size precision gain resistor Rg = 50 kΩ / (G − 1)
Choose Rg so Vref + G × ΔVin(max) stays ≥ 0.2 V inside the INA’s supply rails. Remember Rg’s temperature drift directly becomes gain drift — always use ≤25 ppm/°C metal-film resistors.
3Buffer the REF pin and add differential RFI filter before INA inputs
Drive REF from a unity-gain op-amp (0.5 V for unipolar scales to stay off the 0 V rail, or Vexc/2 for tension/compression). Place symmetric 1 kΩ + 10 nF differential / 1 nF common-mode C0G capacitors right at INA+ / INA− to block mobile-phone RF rectification.
4Use ratiometric excitation (or 6-wire Kelvin sense on long cables)
Share Vexc with the ADC Vref so excitation voltage drift cancels completely. On cable runs > 5 m, use 6-wire load cell cables (Sense+ / Sense−) so excitation cable drop never shrinks sensitivity!
3. Fehlersuche & Multimeter-Diagnose
| Symptom | Ursache | Multimeter-Prüfung | Abhilfe |
|---|---|---|---|
| INA output stays pinned at +4.8 V or 0.05 V regardless of applied weight | Gain too high, open bridge arm, or single-supply negative offset without Vref bias | Measure bridge ΔVin directly with a mV meter and check total bridge resistance (350 Ω) | Raise Rg, bias Vref to ≥0.5 V (or 2.5 V), or replace damaged load cell |
| Output drifts wildly when a phone or VFD operates nearby | RF rectification inside INA input stage due to missing symmetric RFI filter | Check if differential cap C_diff ≥ 10× common-mode cap C_cm right at INA pins | Add matched 1 kΩ resistors + 10 nF diff + 2 × 1 nF CM C0G capacitors next to INA |
4. Häufige technische Fragen (FAQ)
Why should I NEVER drive an instrumentation amplifier’s REF pin with a bare resistor divider?
Any resistance on the REF pin adds directly in series with the internal difference-amplifier feedback network, unbalancing its precision laser-trimmed resistors! Even 100 Ω on REF can degrade CMRR from 100 dB down to 60 dB.
Should I use an INA + MCU ADC or an integrated 24-bit load cell ADC like HX711 / ADS1232?
For slow static weighing (<80 SPS), an integrated 24-bit sigma-delta ADC (ADS1232 / NAU7802 / HX711) with built-in PGA=128 gives dramatically better resolution and lower cost. Choose an analog INA (INA128/INA333) when you need fast dynamic strain/impact bandwidth (>1 kHz) or analog 0–10 V / 4–20 mA output.
Why is 350 Ω preferred over 120 Ω for industrial load cells?
At the same excitation voltage (e.g. 10 V), a 350 Ω bridge draws ~3× less current (28.6 mA vs 83.3 mA), cutting self-heating and reducing sensitivity to lead-wire resistance by nearly 3×.
Why bias Vref to 0.5 V even for a unidirectional compression-only scale?
No single-supply op-amp or INA reaches true 0.000 V linearly, and real load cells often have slight negative zero offset (−0.2 mV). Lifting Vref to 0.5 V keeps zero-load readings solidly inside the linear region for clean digital tare.
What is a 6-wire load cell cable and when is it needed?
A 6-wire cable adds two Sense+ / Sense− feedback wires tied to the bridge excitation nodes so the ADC reference (or excitation regulator) measures the true voltage reaching the bridge — cancelling excitation cable drop on long runs.
How do I check if a Wheatstone bridge load cell is damaged using a multimeter?
Measure Exc+ to Exc− (~350–410 Ω) and Sig+ to Sig− (~350 Ω), then measure each corner diagonal (all four should be equal within ~1 Ω, e.g. ~263 Ω). Finally check insulation resistance from leads to metal body (>500 MΩ).
5. Zugehörige Rechner, Leitfäden & weitere Schaltungen
- Rechner für Dehnungsmessbrücken
- 6-Achs-Kraft-Momenten-Sensor-Rechner (F/T)
- Drehmomentsensor- & Wellen-DMS-Rechner
- Rechner für Instrumentenverstärker-Verstärkung
- Ausgangssignal von Wägezellen
- Rechner für ADC-Auflösung
- NTC-Spannungsteiler + OPV-Impedanzwandler & RC-Filter
- 2-/3-/4-Leiter RTD (PT100 / PT1000) Konstantstrom-Speisung
- Thermoelement-Frontend: CJC, EMV-Filter & Fühlerbruch
- Hochohmiges pH-Glaselektroden-Frontend & Guard Ring
- PNP-Schaltausgang (3-Leiter, Plus-schaltend) — Schaltung & Simulator
- NPN-Schaltausgang (3-Leiter, Minus-schaltend) — Schaltung & Simulator
- NAMUR 2-Leiter Eigensicherheits-Schaltung (IEC 60947-5-6)
- 2-Leiter 4–20 mA Stromschleife & geschützter Empfänger