Calcolatore pirometro IR D:S ed emissività

Diametro spot ottico (D:S) ed errore di temperatura apparente secondo Stefan-Boltzmann.

A 1,5 m con D:S 20:1 lo spot misura 75 mm; un target a 200 °C (ε = 0,85) con ε = 0,95 legge 187,5 °C (−12,5 °C di errore).

Ottica e superficie

Risultato

189,1°C

Apparente 189,1 °C (errore -10,9 °C) · Spot Ø 75 mm

Diametro spot ottico
75,0 mm0,075 m (D:S = 20:1)
Errore emissività
-10,9 °Capparente − reale
Quota riflessa
2,7 %T_bg = 25 °C
ε reale / impostata
0,85 / 0,95opaco ≈ 0,95, acciaio ≈ 0,25

Schema

IR Pyrometer8–14 µm189,1 °C-10,9 °CSpot: 75,0 mmε·T_obj⁴ + (1 − ε)·T_bg⁴
Formule e norme
D_spot = d / (D:S)   T_app⁴ = [ε_true · T_obj⁴ + (1 − ε_true) · T_bg⁴] / ε_set

D_spot = d / (D:S); Stefan-Boltzmann: T_app⁴ = [ε_true·T_obj⁴ + (1 − ε_true)·T_bg⁴] / ε_set (in kelvin).

Il bersaglio deve essere almeno 1,5× più grande dello spot. Metalli lucidi (ε < 0,3) richiedono pirometri a onda corta.

D:S spot size and emissivity error

An infrared pyrometer measures thermal radiation inside its optical cone. At distance d with optical ratio D:S, the target spot diameter is D_spot = d / (D:S). If the instrument emissivity setting ε_set does not match the true surface emissivity ε_true, Stefan-Boltzmann radiation gives:

  • T_app⁴ = [ε_true · T_obj⁴ + (1 − ε_true) · T_bg⁴] / ε_set

Worked example

At 1,5 m with a 20:1 pyrometer, D_spot = 75 mm. Viewing a 200 °C oxidized steel surface (ε_true = 0,85) with ε_set = 0,95 and 25 °C ambient gives an apparent reading of 187,5 °C (−12,5 °C low).

Frequently asked questions

Why do infrared thermometers read low on shiny stainless steel or aluminium?

Polished metals have very low emissivity (ε ≈ 0,08–0,30) and reflect cooler ambient surroundings. Unless ε is lowered or a matte black tape patch (ε ≈ 0,95) is applied, the pyrometer reads far below the true surface temperature.

How big must my target be compared to the optical spot D_spot?

Make sure the hot target is at least 1,5× larger than D_spot so cooler background energy does not spill into the detector field of view.

Can an IR pyrometer measure accurately through glass?

Standard 8–14 µm broadband pyrometers cannot see through ordinary window glass (glass is opaque at 8–14 µm and reflects room temperature). Use a short-wavelength 1,6 µm or 5 µm pyrometer designed for glass/furnace viewing.