SensorCatalog

적외선 온도계 D:S 스폿 & 방사율 오차 계산기

D:S 광학 스폿 직경과 방사율 불일치에 의한 슈테판-볼츠만 온도 오차 계산.

1.5 m·D:S 20:1 스폿 직경은 75 mm이며 200 °C(ε=0.85)를 ε=0.95로 측정하면 187.5 °C(−12.5 °C 오차)로 표시됩니다.

광학 및 표면 조건

결과

189.1°C

지시 189.1 °C (오차 -10.9 °C) · 스폿 Ø 75 mm

광학 스폿 직경
75.0 mm0.075 m (D:S = 20:1)
방사율 측정 오차
-10.9 °C지시값 − 실제값
반사 배경 에너지 비
2.7 %T_bg = 25 °C
실제 / 설정 ε
0.85 / 0.95무광≈0.95, 금속≈0.25

개략도

IR Pyrometer8–14 µm189.1 °C-10.9 °CSpot: 75.0 mmε·T_obj⁴ + (1 − ε)·T_bg⁴
공식 및 규격
D_spot = d / (D:S)   T_app⁴ = [ε_true · T_obj⁴ + (1 − ε_true) · T_bg⁴] / ε_set

D_spot = d / (D:S), T_app⁴ = [ε_true·T_obj⁴ + (1 − ε_true)·T_bg⁴] / ε_set (절대온도 K 기준).

대상 크기는 스폿 직경의 1.5배 이상이어야 하며 광택 금속(ε < 0.3)은 단파장 온도계나 무광 테이프를 사용하세요.

D:S spot size and emissivity error

An infrared pyrometer measures thermal radiation inside its optical cone. At distance d with optical ratio D:S, the target spot diameter is D_spot = d / (D:S). If the instrument emissivity setting ε_set does not match the true surface emissivity ε_true, Stefan-Boltzmann radiation gives:

  • T_app⁴ = [ε_true · T_obj⁴ + (1 − ε_true) · T_bg⁴] / ε_set

Worked example

At 1.5 m with a 20:1 pyrometer, D_spot = 75 mm. Viewing a 200 °C oxidized steel surface (ε_true = 0.85) with ε_set = 0.95 and 25 °C ambient gives an apparent reading of 187.5 °C (−12.5 °C low).

Frequently asked questions

Why do infrared thermometers read low on shiny stainless steel or aluminium?

Polished metals have very low emissivity (ε ≈ 0.08–0.30) and reflect cooler ambient surroundings. Unless ε is lowered or a matte black tape patch (ε ≈ 0.95) is applied, the pyrometer reads far below the true surface temperature.

How big must my target be compared to the optical spot D_spot?

Make sure the hot target is at least 1.5× larger than D_spot so cooler background energy does not spill into the detector field of view.

Can an IR pyrometer measure accurately through glass?

Standard 8–14 µm broadband pyrometers cannot see through ordinary window glass (glass is opaque at 8–14 µm and reflects room temperature). Use a short-wavelength 1.6 µm or 5 µm pyrometer designed for glass/furnace viewing.