PT100 and PT1000 accuracy at 100 °C

What IEC 60751 allows a platinum resistance thermometer at 100 °C, class by class, and what the rest of the measurement adds.

Tolerance by class

ClassTolerance (°C)PT100 (Ω)PT1000 (Ω)
AA±0.27±0.102±1.02
A±0.35±0.133±1.33
B±0.80±0.303±3.03
C±1.60±0.607±6.07

At 100 °C

Resistance
138.506 Ω · 1385.06 ΩPT100 / PT1000
Sensitivity
0.3793 Ω/K · 3.793 Ω/KPT100 / PT1000

Example: the whole measurement

TermAssumptionError
Sensor tolerancePT100, Class A±0.350 K
Lead wires3-wire, 10 m, 0.5 mm²±0.045 K
Transmitter input±0.1 K±0.100 K
Analog output±0.03 % × 200 K±0.060 K
Ambient effect±0.1 % / 10 K × 20 K±0.400 K
Total probable errorRSS±0.55 K
Worst caseΣ±0.96 K

An example with typical datasheet figures, not a statement about any product. Enter your own values in the calculator.

Open this example in the error calculator

How the tolerance is defined

IEC 60751 gives each class as ±(a + b·|t|) °C within its own temperature range: AA 0.1 + 0.0017·|t|, A 0.15 + 0.002·|t|, B 0.3 + 0.005·|t|, C 0.6 + 0.01·|t|. The ranges here are those for wire-wound elements; thin-film elements often have narrower ones, so check the datasheet. The same limits in °C apply to PT100 and PT1000; in ohms the PT1000 figure is ten times larger.