SensorCatalog
ASTM D1293Coaxial BNC + Driven Guard Shield

Frontal Electrodo de Vidrio pH Alta Impedancia y Anillo de Guarda

Nernst ±414 mV (−59.16 mV/pH) · 3.3 V Single Rail + Mid-Rail Vref

Un electrodo pH de 500 MΩ exige un buffer de femtoamperios (LMP7721, 3 fA) y anillo de guarda activo; 1 nA produce un error de 500 mV (>8 pH).

NERNST ACCURATE — E = +177.5 mV → pH 4.00

At 25 °C (slope -59.16 mV/pH), 3 fA bias × 500 MΩ produces only 1.50 µV error; Guard Ring locks PCB leakage to 0 V.

Flujo normalAlerta / LimitaciónFallo / SobrecorrienteVelocidad ∝ |I|
pH Glass ProbepH 4.0+177.5 mVR_glass = 500 MΩRef: 1.65 VDriven Guard Ring (0 V Drop)Vref = 1.65 V+−LMP77213 fABuffered pH Out1.83 VV_out = Vref + E_pHNernst −59.16 mV/pH⚠ 1 nA Ib × 500 MΩ = 500 mV ERROR!

Nernst Electrode EMF

+177.5 mV

Shifted ADC Output

1.827 V

Measurement pH Error

-0.00 pH

Temp-Comp Nernst Slope

-59.16 mV/pH

Driven Guard RingActive @ 1.83 V
Op-Amp Input Bias Ib3 fA (LMP7721)
Reported pHpH 4.00
pH

pH 7.00 = 0 mV; acidic (<7) is positive mV, alkaline (>7) is negative mV.

°C

Nernst slope is −59.16 mV/pH at 25 °C and scales with Kelvin T.

MΩ

Typically 100–500 MΩ at 25 °C (rises >1 GΩ in cold water!).

V

Shifts ±414 mV pH probe voltage cleanly above 0 V for single-supply ADC.

Preajustes:

Fórmulas paso a paso en vivo

Temperature-Compensated Nernst Equation

E = −59.16 mV · [(T + 273.15) / 298.15] · (pH − 7) = -59.16 mV/pH × (4.0 − 7.0)

⇒ +177.48 mV

Op-Amp Input Bias Offset Voltage

V_err = I_b × R_glass = 3 fA × 500 MΩ

⇒ +1.50 µV

Lista de materiales dinámica (BOM)

Ref.ComponenteEspecificación calculadaEstado
U1LMP7721 / LMC6001 Femtoamp Op-AmpIb = 3 fA typ, dedicated guard pinsOK
J1PTFE-Insulated BNC + Driven Guard RingNo solder mask over guard ringOK

1. Descripción general y principio de funcionamiento

A pH glass electrode is one of the highest-impedance sensors in engineering: 100 MΩ to 1 GΩ source resistance generating ±414 mV via the Nernst equation (−59.16 mV/pH at 25 °C). Measuring it accurately requires a femtoamp-bias op-amp (LMP7721 / LMC6001), a driven Guard Ring around the input pin, and a precision mid-rail Vref shift.

Why Ordinary Op-Amps Fail Catastrophically: Even a "good" 1 nA input-bias op-amp flowing through a 500 MΩ glass bulb creates 1 nA × 500 MΩ = 500 mV error — more than 8 full pH units! Only fA-class amplifiers (Ib = 3 fA → 1.5 µV error) preserve ±0.01 pH accuracy.

Driven Guard Ring Physics: Surrounding the high-Z BNC center pin with a bare copper ring driven by the unity-gain buffer output keeps the potential difference across the FR4 surface at 0.000 V, reducing PCB moisture leakage current to zero!

Frontal Electrodo de Vidrio pH Alta Impedancia y Anillo de Guarda — SensorCatalog
Electrodo pH y GuardaPNG 1200×630

2. Step-by-step: Building a ±0.01 pH glass-electrode front-end

  1. 1Select a femtoamp-class unity buffer (LMP7721 / LMC6001)

    LMP7721 provides dedicated empty guard pins right next to the input pin for effortless PCB guard routing.

  2. 2Bias the reference electrode to a clean 1.65 V / 2.048 V Vref

    Shifts the ±414 mV Nernst swing cleanly into 1.2–2.1 V on a single 3.3 V rail.

  3. 3Route an exposed-copper driven Guard Ring on top and bottom layers

    Remove solder mask over the guard ring and ultrasonically wash all rosin flux residue.

  4. 4Measure solution temperature (PT1000/NTC) for Nernst slope ATC

    Slope shifts from −54.2 mV/pH at 0 °C to −64.1 mV/pH at 50 °C — ATC is mandatory outside 25 °C.

3. Diagnóstico en campo y pruebas con multímetro

SíntomaCausa raízPrueba con multímetroSolución
pH reading jumps wildly whenever a hand approaches the boardUnshielded high-impedance BNC/trace picking up 50/60 Hz capacitive body humPlace grounded metal shield over front-end and check stabilityUse PTFE coaxial BNC, driven Guard Ring, and metal EMI shield can

4. Preguntas frecuentes (FAQ)

Why is the pH electrode voltage 0 mV at pH 7.00?

Because the internal reference buffer inside the glass bulb is formulated at pH 7.00, so zero hydrogen-ion concentration gradient exists across the glass membrane at neutral pH 7.

5. Calculadoras relacionadas, guías y otros circuitos

↑ ↓ elegir · Intro abrir · Esc cerrar